The Purpose of the PV Shift is to make the temp controller as accurate as you can to the plate temperature.

Introduction

Omron E5CVS PV Shift E5CSVJPG.JPG
Utiliser des vêtements de protection pour les mains
Avertissement pour surface chaude

Étape 1 - Unit Operation

Using the Mode switch step through until you reach


Omron E5CVS PV Shift Mode key.jpg



Étape 2 - Checking if Input shift is ON.

Omron E5CVS PV Shift Setting input shift.jpg



Étape 3 - Adjust Shift value until Temperature at Plate is same as on Display

H0 = No offset. Temp at Plate =230 degrees - Display on E5CSV = 240 Degrees

L10 = - 10 degree. Temp at plate = 230 degrees - Display E5CSV = 230 Degrees

H10 = +10 degree . Temp at plate = 230 degrees - Display E5CSV = 250 Degrees

Étape 4 - Checking Protection button is not locked. Dip Switch P

Omron E5CVS PV Shift Protect switch key.jpg